Quick answer: CV regulates voltage until the load reaches the selected current limit. CC regulates current by adjusting voltage within the selected ceiling. A CV-to-CC transition may be expected operation, not a malfunction.

On this page
  1. Constant voltage operation
  2. Constant current operation
  3. Calculate an expected transition
  4. CC is different from protection shutdown
  5. Applications and limits
\"CV

Constant voltage operation

In CV mode the supply controls voltage at the active sensing point. The load determines current as long as demand stays below the selected current ceiling and the instrument’s operating limits. Lead resistance can create a different voltage at the DUT if sensing is local.

Constant current operation

In CC mode the supply adjusts voltage to hold the selected current, up to the voltage ceiling. If the load changes, the voltage needed to sustain that current changes too. A source may be unable to establish the requested operating point when the load or power limit prevents it.

Calculate an expected transition

For an ideal resistor, I = V/R. With 10V selected and a 1A current limit, a 20Ω resistor draws 0.5A in CV mode. A 5Ω resistor would demand 2A, so the ideal CC operating point is 5V at 1A. Real startup or electronic loads may not behave as fixed resistors.

CC is different from protection shutdown

A CV-to-CC transition may be ordinary operation. OCP or other fault functions can instead disable output or latch an error. Read the instrument’s documented behaviour before treating every low-voltage indication as a fault.

Applications and limits

CV/CC control supports many device tests and suitable sourcing tasks. Batteries require a validated chemistry-specific protocol and external-energy compatibility. Discharge, regenerative operation or dedicated photovoltaic/battery simulation requires documented equipment capabilities beyond generic CV/CC operation.

Practical checks

CheckWhat to verify
Resistive loadCheck the expected current using I = V/R.
Current ceilingChoose a limit appropriate for the DUT and startup behaviour.
Voltage collapseIdentify CC operation before assuming a defective supply.
OCPCheck whether OCP latches off or has a different reset behaviour.

Compare a supply for your test

Explore variable DC power supplies or review bench DC power supplies. Specify your operating voltage, current, power, load type and required controls for selection a configuration.